This research integrated a MEMS electrostatic driven micro-probe and a laser Doppler vibrometer for non-contact vibration mode scanning probe microscope system design. The microprobe tip was placed in perpendicular to the sample surface, and the built-in capacitor on the microprobe was excited to vibrate by a sinusoidal drive voltage to generate Coulomb electrostatic force. The applied frequency is right at the structure natural resonant frequency of the microprobe. Then let the sample carried by a Z-stage move up. When the sample gets closer to the microprobe, the Van Der Waal’s force between the sample and microprobe would become larger, and the microprobe vibration amplitude would be reduced, and which can be determined by a laser Doppler vibrometer. Since the probe vibration amplitude is proportion to the distance between the probe tip and the sample surface. Thus one can detect the sample surface profile, by moving the probe tip at a constant height, and using a laser Doppler vibrometer to obtain the topography with the amplitudes of microprobe vibration history. The accuracy of the proposed system is about 10 nanometers with a gauge meter.


Keywords: Microprobe, Non-Contact Vibration Mode, Scanning Probe Microscope


REFERENCES


  1. [1] Feynman, R. P. “There’s Plenty of Room at the Bottom,” American Physics Annual Conference (1959), also appeared at Caltech’s Engineering and Science, February (1960).
  2. [2] Binnig, G., Roher, H., Gerber, H. and Weibel, E. “Surface Studied by Scanning Tunneling Microscopy,” Physical Review Letters, Vol. 49, pp. 5761 (1982).
  3. [3] Neubauer, G., “Force Microscopy with a Bidirectional Capacitor Sensor,” Rev. Science Instrument, Vol. 61, pp. 22962308 (1990).
  4. [4] Binnig, G., Quate, C. F. and Gerber, C. H., “Atomic Force Microscope,” Physical Review Letters, Vol. 56, pp. 930933 (1986).
  5. [5] Fung, R. F. and Huang, S. C., “Dynamic Modeling and Vibration Analysis of the Atomic Force Microscope,” ASME, Journal of Vibration and Acoustics, Vol. 123, pp. 502509 (2001).
  6. [6] Seeger, J. I. and Crary, S. B., “Stabilization of ElectroStatistically Actuated Mechanical Devices,” Transducers ’97, Chicago, IL, p. 1133 (1997).
  7. [7] Lee, S. Y. and Lin, S. M., “Exact Solutions for the Rations of an Elastically Restrained Non-Uniform Timoshenko Beam with Tip Mass,” AIAA Journal, Vol. 30, pp. 29302934 (1992).
  8. [8] Holscher, H., Schwarz, U. D. and Wiesendanger, R., “Calculation of the Frequency Shift in Dynamic Force Microscopy,” Applied Surface Science, Vol. 140, pp. 344351 (1999).
  9. [9] Sasaki, N., Tsukada, M., Tamura, R., Abe, K. and Sato, N., “Dynamics of the Cantilever in Non-Contact Atomic Force Microscopy,” Applied physics A: Materials, Vol. 66, pp. 287291 (1998).
  10. [10] Rabe, U., Janser, K. and Amold, W., “Vibrations of Free and Surface-Coupled Force Microscope Cantilevers: Theory and Experiment,” American Institute of Physics, Vol. 67, pp. 32813293 (1996).
  11. [11] Fung, R. F. and Huang, S. C., “Dynamic Modeling and Vibration Analysis of the Atomic Force Microscope,” ASME, Journal of Vibration and Acoustics, Vol. 123, pp. 502509 (2001).